Sample Type Required: Deparaffinized and re-hydrated tissue sections on positively charged slides.

Fixative: 10% Neutral Buffered Formalin

Step

Reagent

Time

Technical Notes

1

Periodic Acid

0.5% Solution

5 Minutes


2

Distilled Water

3 Changes


3

Schiff Reagent

15 Minutes

Gluteraldehyde should be avoided as a tissue fixative since it is a dialdehyde that will react with this reagent and give false-positive PAS staining.

4

Potassium Metabisulfite

0.55% Solution

1 Minute in 2 Changes

While some technicians omit this reagent step in the protocol with no problems, metabisulfite rinses may be necessary to remove any excess leucofuchsin left over after exposure to the Schiff reagent. Highly chlorinated water can cause these excess molecules to nonspecifically stain other tissue elements in the section.

5

Running Tap Water

Up to 10 Minutes

Warm to hot running water develops the PAS stain and gives a more intense staining reaction than cold water.

Expected Results
The following elements will show positive PAS (Rose Red) staining:

  • Glycogen
  • Neutral mucins
  • Some epithelial mucins
  • Basement membranes
  • Fungal walls

Post Staining Procedure: Tissue sections should be rinsed well in distilled water, dehydrated with 95% and absolute alcohols, cleared, and cover-slipped.